Technical parameters
Test item: consistency of K + , Na + , Cl - , Ca 2+ etc. ions and PH value.
Various models measure projects as follow:
Type A: Measuring electrolyte ions K + , Na + , Cl - .
Type B: Measuring electrolyte ions K + , Na + , Cl - ,iCa 2+ .
Type C: Measuring electrolyte ions K + ,Na + ,Cl - ,iCa 2+ ,pH value.
Type D: Measuring electrolyte ions K + ,Na + ,Cl - ,Li + .
| Sample | Serum, urine etc |
| Sample dosage | 100ul-200ul; adjustable |
| Test method | Ion selectivity electrode technology |
| Test speed | ≤60 samples/ hour |
| Memory | 2000 tests |
| Display | 320*240 LCD |
| Inner printer | 24 bits thermal printer |
| Output | RS-232C tie |
| Store condition | Temp: 0—40°C; Relative Humidity: ≤85% |
| Power supply | AC90-264V;47-440Hz |
| Input efficiency | ≤ 80 VA |
| Electric shock proof level | Level 1; Overvoltage category: Class I; Pollution level: Level 2; Management category: Class II; |
| Weight | 5Kg |
| Classified according to working system | continuous operation instrument |
| Property | IVD in vitro diagnostic equipment |
The Picture of Touch Screen Electrolyte Analyzer



Structure Introduction
Product performance,structure and composition: This product is composed of main
machine and ancillary reagent.
Front view:

1: Touch Screen 2: Electrode Observation Window
3: Suck Sample Needle 4: Front Baffle
5: Sample Pool

1: Electrode 2: Point Liquid Valve
3: Peristaltic Pump 4: Reagent Storehouse
5: Printer
Right view

1. Soft tube 2. Waste liquid tube 3. Reagent
Back view

1: Cooling Fan; 2: COM RS-232 1;
3: COM RS-232 2; 4: Equipotential wiring terminal;
5: Power input,1 00—240VAC;50—60Hz;
6: Fuse; 7: Power switch;